Catalog Number DYND11003H
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/20/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "product consistently leaking, possibly from metal clamp on drainage bags".
 
Additional Manufacturer Narrative
The customer/reporter described product consistently leaking, possibly from metal clamp on drainage bags, and stated it appears to be the seal on the bottom of the drainage bag that is leaking.Medline quality did not receive a sample of the reported product; medline quality did receive a video, reviewed the inspection records, and reviewed the provided video, which revealed leakage around the seal of the slide-tap drainage valve at the bottom of the bag.The reported issue was verified by assessment of the provided video.The likely root cause is external damage or a seal defect during manufacturing; there are currently no known trends, and medline quality will continue to monitor for the issue through weekly complaint meetings and monthly trending analysis.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
TRAY,FOLEY,CATH,SIL-ELAST,16FR,10ML,BG
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
nathan bersch
three lakes drive
northfield, IL 60093--275
MDR Report Key50856416
Report Number1417592-2026-47882
Device Sequence Number21220814
Product Code OHR
UDI-Device Identifier10080196991471
UDI-Public10080196991471
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYND11003H
Device Lot Number96923080002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2023
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-