| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable creatine kinase results for 12 patient samples on a cobas 4000 c311 stand alone system.As a representative example, three patient results are provided.Patient 1's initial result was 680 u/l.The repeat result from a different c311 was 66 u/l.Patient 2's initial result was 1008 u/l.The repeat result from a different c311 was 97 u/l.Patient 3's initial result was 1890 u/l.The repeat result from a different c311 was 197 u/l.The repeat results were deemed to be correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The cobas 4000 c311 stand alone system serial number is (b)(6).The calibration failed 3 times with std.E flags and only passed when using a new reagent pack.The std.E flag indicates issues with calibrator assignment or preparation; calibrations with the new reagent cassette were overall inconspicuous but had slightly higher values for both standards.As calibration was successful and correct results were acquired using the same reagent pack on the new device, a reagent/reagent pack-specific issue can be excluded as the root cause.The alarm trace does not show any relevant alarms for the time around the measurements.The investigation is ongoing.
|
| |
|