It was reported that: "we have had intermittent issues with not being able to flush.Today, 6 of the sets in a row would not flush.We tried different ns flushes, and they would not flush.The clamp was not engaged, and the luer lock was secure.".
|
|
The account reported finding unable to flush extension set; the customer reported that 6 of the sets in a row would not flush, different ns flushes were tried, the clamp was not engaged, and the luer lock was secure.A sample was received for investigation; to evaluate the sample, medline connected a syringe to the device and pressed on the plunger, and at first the syringe plunger would not move and medline was unable to flush the device, however after playing around with the device for a bit, medline was able to flush the extension set.Based on the complaint details, sample evaluation, and investigation, the complaint is confirmed.Medline was unable to determine the exact root cause for why the device could not be flushed initially but later could be flushed, and the complaint was provided to the manufacturing facility for further evaluation while the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|