Medline Industries, LP MEDLINE; SET,IRRIGATION,CYSTO,Y-TYPE,90 Back to Search Results
Catalog Number DYND19122
Medical Device Problem Codes Disconnection (1171); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/03/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "they leak at the connection.".
 
Additional Manufacturer Narrative
The customer reported the affected irrigation sets "leak at the connection." urology qa launched an investigation with product testing, placing the product on hold, testing at the dvv lab to confirm discrepancies between old and new lots, and validating a replacement vendor.After extensive testing, the dvv lab was able to confirm failures in the associated lots from customer complaints; the vendor confirmed that the tubing issue was the result of a compromised batch of raw material from their supplier.Urology qa opted to freeze all sales and move to the new vendor, and stated that the disconnections were determined to not pose immediate health hazards to the patient and that repeated regulatory investigations did not identify any direct patient harm nor medical intervention.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
SET,IRRIGATION,CYSTO,Y-TYPE,90
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
zelek alexander
three lakes drive
northfield, IL 60093--275
MDR Report Key50856545
Report Number1417592-2026-47915
Device Sequence Number21206853
Product Code LJH
UDI-Device Identifier40080196993827
UDI-Public40080196993827
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K970946
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYND19122
Device Lot Number18322060001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/03/2023
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-