The customer reported that the bp was "not giving them the proper reading," was "giving them error," and one error "is error 4." medline received 1ea wrist bp monitor in used condition; the sample came with 2 aaa batteries, 2 aaa control batteries were inserted and the sample was tested according to the instructions for use, and the sample turned on and would inflate the cuff; however, the monitor would then display "er 4." medline changed the cuff and then attempted to perform several consecutive blood pressure measurements, but the issue continued; the complaint was confirmed, and the issue shown via investigation was caused due to a functional malfunction, not allowing the monitor to inflate the cuff properly.Cause was documented as "cannot determine cause w/received sample," and the sample did not operate as intended.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|