The account reported two manifolds broke during their ventrical injection with their medrad injector, and the reported issue was ra separated from manifold.A photo and physical sample were received to aid the investigation; finished good production records and finished good pack build were reviewed, with no instances of line downs, run withouts, deviations, non-conformance reports, or non-conformances noted that would have contributed to the reported issue.The same manifold was observed with the ra separate from the body, confirming the customer's report.The root cause of the rotating adaptor separating from the manifold body was determined to be an inadequate swage during the assembly process of the manifold, and the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|