Olympus Winter & Ibe GmbH Video telescope "EndoEYE", 10 mm, 0°, NTSC, HD compatible, autoclavable; Laparoscopes, Video
|
Back to Search Results |
|
| Model Number |
WA50011A |
| Medical Device Problem Codes |
Break (1069); Poor Quality Image (1408); Dent in Material (2526); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported a black shadow in the upper right corner and side of the viewing area involving an olympus laparoscopes, video.Additionally, during the initial evaluation of the device, fiber breakage, dents, and scratches on the distal frame, and dents and scratches on the outer tube were observed.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The date of the event is unknown.Additional information will be provided once available.The final investigation has been completed.The failure reported by the customer, as well as the issues identified during inspection, were confirmed.Based on the results of the investigation a root cause could not be identified.Olympus will continue to monitor the field performance of this device and evaluate any additional information received through post-market surveillance activities.
|
| |
|
Search Alerts/Recalls
|
|
|