The reported issue was readings high on all patients; when compared to other bp machines, the patient is then in a normal range for bp.Medline received 2 ea digital wrist blood pressure monitors; upon visual inspection there was no cosmetic dammage, and during functional testing with new aaa batteries the monitors responded to all button inputs correctly and responded to simulating blood pressure at 120/80 @ 80 bpm and 150/100 @ 80 bpm.The following values were noted: monitor #1: 117/80 @ 79 bpm and 152/98 @ 79 bpm, monitor #2: 117/81 @ 80 bpm and 154/99 @ 80 bpm; these are within acceptable ranges, and the complaint was not confirmed.The cause was unconfirmed defect - not confirmed with received sample; there are currently no known trends for this product, and no further action was taken.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|