| Model Number |
PCF-H190I |
| Medical Device Problem Codes |
Corroded (1131); Poor Quality Image (1408); Communication or Transmission Problem (2896); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the colonovideoscope for a separate issue.During the initial evaluation of the returned device, a b30 (scope communication error), noisy image, corrosion on the connecting tube, and no illumination were observed.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the final investigation of the device the initial evaluation finding were confirmed.A root cause for the reportable malfunction, scope communication error (b30), identified during the device evaluation was traced to corrosion of the plug unit.The most likely cause of the noisy image was damage to the charge-coupled device (ccd) unit.The most likely cause of the no illumination condition was breakage of the light guide bundle.Additionally, the corrosion on the connecting tube was traced to component failure.Olympus will continue to monitor the field performance of this device.Should any additional relevant information become available, a supplemental report will be submitted.
|
| |
|