| Model Number |
N/A |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Corneal Perforation (1792)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Other
|
|
Event or Problem Description
|
|
On 09/01/2026, doctor (b)(6) reported to cas that on (b)(6) 2026, they experienced a perforated ak during procedure id #(b)(6).Site is seeking review of the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Review of procedure id #(b)(6) video and imaging shows patient eye movement through-out the procedure.The second arcuate incision at 228 shows a possible posterior perforation with bubbles on the posterior surface of the cornea.It is recommended lowering the arcuate depth to 80% from 85% to reduce this risk if the surgeon is agreeable.Also confirming the patient is locked to the pid arm and not moving is recommended.System functioned as designed.No further clinical follow - up required.
|
| |
|