Synthes GmbH REDUCTION FORCEPS WITH SERRATED JAW-RATCHET 144MM Back to Search Results
Catalog Number 399.99
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the connector screw is loose.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: h3, h6: investigation summary the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found the weld from the connector pin was broken.Due to breakage the ratcheting mechanism does not work as intended as the handle is loose.The observed condition of the device was consistent with a component failure that was caused by exposure to unintended forces.Properly handling and attention to the approved use of the device diminishes the risk of failure.A dimensional inspection was unable to be performed due to the post-manufacturing damage.The overall complaint was confirmed as the observed condition of the red forceps serrated jaw-ratchet 144 would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot part: 399.99 lot: a7ga05 product code: n/a date of manufacture: week 05, 1997 manufacturing location: tuttlingen part expiration date: n/a nonconformance noted: n/a a manufacturing records evaluation is not possible, the dhr is no longer available due to the age of the instrument (over 29 years old).Therefore the exact date of manufacture is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REDUCTION FORCEPS WITH SERRATED JAW-RATCHET 144MM
Common Device Name
Forceps
Manufacturer (Section D)
Synthes GmbH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
Synthes Tuttlingen
unter hasslen 5
tuttlingen 78532
GM   78532
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
+1(908)386-3534
MDR Report Key50857561
Report Number8030965-2026-09642
Device Sequence Number13136035
Product Code HTD
UDI-Device Identifier10886982203175
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
Exempt
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number399.99
Device Lot NumberA7 GA 05
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/27/2026
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 09/08/2026
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-