The reported issue was adhesive issue dressing, with the backing on the new suresite dressing not pulling off properly and sometimes tearing the whole dressing.Per the investigation of the samples received, unused kits were returned for evaluation; dressings were removed from the kit, peeled from the release liner, and then applied to a laminate surface, and finished good production records were reviewed with no nonconformances noted by production or qc personnel.All of the dressings performed as expected, and the sample evaluation was unable to reproduce the reported issue.A root cause cannot be determined, and there has been no change to the reportability decision associated with this complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|