| Model Number |
LifeVest WCD 5100 System |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor returned a monitor and reported monitor was displaying a service code and adjust belt messages.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of monitor has been completed.The reported problem (service codes, activity report, and adjust belt messages) has been confirmed.The cause for the failure was isolated a shorted u604 component on the computer/analog pca.The root cause for the shorted u604 component could not be positively identified.There was no adverse event that resulted from the damaged monitor.
|
| |
|