| Model Number |
71953-01 |
| Medical Device Problem Code |
Product Quality Problem (1506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported their adc device is too hot to hold.There was no report of fire, smoke, explosion, or shock.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The reader was returned and visual inspection has been performed since the reader had been de-cased prior to assessment.This issue is unable to be tested due to the product not being in a condition that allows the re-testing of the product to reproduce or verify the customer¿s complaint.However, the reader no longer turns on and battery was observed to be swollen.This is unable to test.At the time of complaint intake, this event was deemed not reportable and troubleshooting determined that no further investigation of the device was required.This complaint has been reassessed as reportable upon retrospective review associated with a capa and is being submitted with all available information.Section h6 (investigation findings) code c20 (no findings available) and d15 cause not established was selected, as adc was unable to perform further testing on the returned device.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|