• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 700 LGX? TENACIO? Pump; prosthesis, penis, inflatable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 700 LGX? TENACIO? Pump; prosthesis, penis, inflatable Back to Search Results
Model Number 72404452
Medical Device Problem Codes Inflation Problem (1310); Mechanical Problem (1384); Nonstandard Device (1420)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient with an inflatable penile prosthesis (ipp) presented with difficulty inflating the device, unable to easily cycle the device, and described as being hard to squeeze with slow deflation.The patient reported needed to pump the device multiple times with minimal inflation achieved.During physician consultation, the physician confirmed that the device was not inflating as intended.The patient was guided on proper pumping technique, including anchoring and holding the pump by tubing and squeezing from the bottom of the bulb to obtain more effective compressions.A surgical procedure was performed during which the pump was explanted and replaced.There were no patient complications.
 
Additional Manufacturer Narrative
Model number/catalog number: 720185-01.Serial number: n/a.Batch/lot number: 1100880320.Model/catalog description: reservoir flat iz 100 ml unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.The reported device performance allegation cannot be confirmed.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the event of mechanical malfunction (leaks, incomplete inflation/ deflation, kinking) was defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available, a conclusion code of cause not established was assigned to this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700 LGX? TENACIO? Pump
Common Device Name
prosthesis, penis, inflatable
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50857745
Report Number2124215-2026-50483
Device Sequence Number21245511
Product Code JCW
UDI-Device Identifier00191506021436
UDI-Public(01)00191506021436(17)270927(10)1100801437
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number72404452
Device Catalogue Number72404452
Device Lot Number1100801437
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2026
Initial Report FDA Received Date09/22/2026
Date Device Manufactured09/27/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age62 YR
Patient SexMale
-
-