| Model Number |
470500-14 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted partial nephrectomy procedure, a rubber piece from a universal seal tore off and fell inside the patient.The customer confirmed that the fragment was retrieved and removed from the patient's abdomen during the same robotic procedure.No patient complications were reported as a result of the event.The procedure was completed as planned.Intuitive surgical, inc.(isi) has attempted to gather additional information regarding the reported event; however, no response has been received.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the universal seal to perform failure analysis.The universal seal was analyzed and found to broken rubber gasket, the universal seal was found to have a broken rubber piece measuring approximately 0.16 in × 0.07 in.The broken piece was returned with the accessory, and no missing pieces were identified during visual inspection.No additional damage was observed on the accessory.
|
| |
|