| Catalog Number |
05.001.201 |
| Medical Device Problem Codes |
Physical Resistance/Sticking (4012); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the battery handpiece device had a sticky trigger, failed the leak tightness test, the k-wire could not be inserted, the device would not hold/secure the battery, the housing was cracked/damaged, and the device had component damage.It was further determined that the device failed pretest for general condition, leakage test, check the cannulation of the handpiece, check for sticky triggers, check the roundness of the housing, and check falling out protection.It was noted in the service order that prior to an unspecified surgical procedure it was observed that the device did not work anymore.This event did not occur during a surgical procedure.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H11: additional narrative:
device evaluation: the actual device was returned for evaluation.The battery handpiece device was evaluated and the reported condition that the device was not working anymore was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the device had a sticky trigger.The assignable root cause of this condition was determined to be traced to component failure due to wear.
|
| |
|
Search Alerts/Recalls
|
|
|