Product complaint #(b)(4)
this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative:
h3, h6: investigation summary
the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found threads from the tip end were stripped.No other issue was identified over the surface of the device.The observed condition of the device was consistent with a component failure that was caused by exposure to unintended forces like assembly of the device in a misaligned position.Properly handling and attention to the approved use of the device diminishes the risk of failure.A functional test was unable to be performed due to the post-manufacturing damage.A dimensional inspection was unable to be performed due to the post-manufacturing damage.The overall complaint was confirmed as the observed condition of the extractscr f/tib+fem nails would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to a component failure, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review
part: 03.010.000-us
lot: 87p2877
manufacturing site: hägendorf
release to warehouse: 28-january-2021
a dhr evaluation was performed for the finished device article 03.010.000-us lot 87p2877, and no non-conformances were identified.
|