| Catalog Number |
03-5300-3 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
08/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A user facility¿s licensed practical nurse (lpn) reported that a 5008x bloodline leaked fifteen minutes after the initiation of a patient¿s hemodialysis (hd) treatment.The machine, a fme 5008x machine, did not alarm, but was not expected to.A fme dialyzer was in use.The leak originated from the arterial disc.The patient¿s estimated blood loss (ebl) was approximately 150ml.The patient was restarted on a new machine and treatment completed successfully with new supplies.There was no patient injury or medical intervention required due to the reported event.The complaint device is not available to be returned to the manufacturer for evaluation.
|
| |
|
Event or Problem Description
|
|
A user facility¿s licensed practical nurse (lpn) reported that a 5008x bloodline leaked fifteen minutes after the initiation of a patient¿s hemodialysis (hd) treatment.The machine, a fme 5008x machine, did not alarm, but was not expected to.A fme dialyzer was in use.The leak originated from the arterial disc.The patient¿s estimated blood loss (ebl) was approximately 150ml.The patient was restarted on a new machine and treatment completed successfully with new supplies.There was no patient injury or medical intervention required due to the reported event.The complaint device is not available to be returned to the manufacturer for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
|
| |
|
Search Alerts/Recalls
|