• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; LIFT,PATIENT,STAND ASSIST,ELECTRC,40

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; LIFT,PATIENT,STAND ASSIST,ELECTRC,40 Back to Search Results
Catalog Number MDS400SAB
Medical Device Problem Codes Battery Problem (2885); Electrical Shorting (2926)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the battery is being charged overight but they're getting a low battery reading," "lift has no function, is beeping and displaying low battery," and "the cord has a short.".
 
Additional Manufacturer Narrative
The customer stated that "the battery is being charged overight but they're getting a low battery reading," and per customer report, "lift has no function, is beeping and displaying low battery"; the customer later advised that "the cord was being unplugged wrong.So the cord has a short." the investigation documented that the sample and/or photo was not available, pcn history was reviewed with no product changes identified that could have contributed to the alleged product failure, and quality control inspection records could not be found for the serial number at that time.The investigation results were not confirmed, and the updated cause code was unconfirmed defect, sample required to determine cause.Based on the review of the details provided for the complaint investigation, medline was unable to confirm the complaint issue or that the product failed to meet specifications, and if a sample, photo and/or additional information becomes available this complaint will be reopened to be further investigated.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
LIFT,PATIENT,STAND ASSIST,ELECTRC,40
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
maya mcknight
three lakes drive
northfield, IL 60093--275
MDR Report Key50858420
Report Number1417592-2026-48101
Device Sequence Number18648359
Product Code FSA
UDI-Device Identifier10193489019254
UDI-Public10193489019254
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS400SAB
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2024
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-