|
The account reported finding "little to no suction," and the complaint transcript noted "low suction at high setting, does not suction otherwise." no sample or photo was received for the investigation, and trending was reviewed.With the information received and no sample to evaluate, medline was not able to confirm what may have caused the issue; the cause was documented as "sample required to determine cause." our quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|