| Model Number |
71003 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported during an implant procedure there was an unknown problem inserting the lead into the ipg.A second lead was attempted and eventually successful.High impedances were being seen with both leads.Procedure was eventually successful.No reports were received of significant delay to procedure and no harm to patient or staff.
|
| |
|
Additional Manufacturer Narrative
|
|
After conducting a post-acquisition, retrospective review of nalu peripheral nerve stimulation complaint records, boston scientific determined that this event meets mdr reporting criteria.This incident is being re-opened/updated per pcar42 remediation activities.The device evaluation was unable to be completed.The devices were confirmed and receipt of decontaminated lead and ipg.Subsequently the ipg was inadvertently discarded.
|
| |
|