| Model Number |
360-1400P |
| Medical Device Problem Code |
Protective Measures Problem (3015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter stated that the pump had fluid ingress and the auxiliary alarm did not operate as expected.They stated that it failed to alarm.Mmd did follow up with the initial reporter, who advised that no patient had been involved and that the complaint had occurred during testing.No additional information was provided."[complaint- (b)(4)]".
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to mmd for evaluation.A dhr review was completed and no non-conformances were found.When the device was returned to mmd for investigation, the bt2 battery and pcb board had been damaged by fluid ingress and failed, causing the auxiliary alarm to fail.The pump was serviced to correct the issue.
|
| |
|