The customer reported gloves are tearing at cuff.Samples were evaluated by visual inspection, physical properties test, thickness measurement, and donning test.The complaint was confirmed based on the images, and medline was able to replicate tearing at the cuff; one returned glove was observed with torn defect at the cuff area, with no pinhole found by the vendor.The cause was updated to dc: unknown responsible party - cannot determine cause w/received sample; based on the investigation conducted on the dhr and tests conducted on the returned and retained samples, no abnormality was found that could lead to glove tearing or breaking, and the gloves conformed to the quality requirement.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|