| Model Number |
IntelliVue Patient Monitor MX700 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Philips received a complaint on an intellivue mx700 patient monitor indicating a discrepancy in spo2 readings.The device was not in clinical use at time of event; there was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation was performed by an authorized philips biomedical engineer (bme) who was onsite.Stated the monitor and attached spo2 module were tested using a calibrated fluke prosim 8 simulator and accurately displayed all tested saturation values.Based on the results of the information available, it was determined that the product was performing per specification and there was no product malfunction and the most likely cause of the reported problem is unknown.The reported issue could not be reproduced therefore, the cause of the event could not be established.The bme stated the device was working to device specifications.
|
| |
|