| Catalog Number |
TRG300L |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that "the glove cuffs easily ripped, gloves came out split in too and they ripped while wiping patients.".
|
| |
|
Additional Manufacturer Narrative
|
|
The customer reported that the glove cuffs easily ripped, gloves came out split in two, and they ripped while wiping patients.Medline quality received 1 bx of the reported product; a torn glove was included, confirming the complaint.A functional assessment of the gloves was conducted by performing a donning test.The gloves were donned without issue, no abnormalities or defects were found, and the reported issue of tearing could not be replicated.A root cause could not be determined based on inspection of the samples.There are no known trends for this product lot.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This mdr is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|
| |
|