Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative:
corrected: h3
h3:h6
the product was returned to depuy synthes for evaluation, and an investigation was conducted in accordance with depuy synthes procedures.Visual inspection of the returned device found the tip was broken apart.Broken fragment was returned attached to mating device insert-handle radioluc fem recon nail.No other anomalies were noted.Surgical technique was reviewed and the following relevant statements were found:
to use a hammer, screw the driving cap into the insertion handle and tighten with the ratchet wrench.While applying light blows, monitor the tip of the nail using image intensification.Verify that there is no evidence of impingement distally.Remove the driving cap once the nail has been seated.Notes: using light blows, the hammer can also be used with the driving cap to back slap the nail if the nail has been slightly over inserted.Alternatively, the hammer guide can be attached to the driving cap to facilitate the use of the hammer.Confirm that the driving cap is tightly connected to the insertion handle, as hammering may loosen the connection.Precautions:
do not strike the insertion handle directly.The reported complaint was confirmed based on the evaluation results.The investigation traced the potential cause to a component failure that was caused by exposure to unintended forces.Properly handling and attention to the approved use of the device diminishes the risk of failure.It has been determined that no corrective or preventive actions are proposed.There is no indication that a design or manufacturing issue caused the reported condition.As part of depuy synthes quality processes, all devices are manufactured, inspected, and released in accordance with approved specifications.Additional monitoring for potential safety signals is conducted through complaint trend analysis as part of post-market safety surveillance activities.Device history review = > a manufacturing record evaluation was performed for the finished device
product code: 03.010.523
lot number: 21p8078
it was electronically reviewed and no nonconformances / manufacturing irregularities
were identified during the manufacturing process.The product was released on: 17 feb 2020
manufacturing site: jabil bettlach
dhr review retrieved from (b)(4)as the impacted products share the same lot number.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|