| Model Number |
LifeVest WCD 4000 System |
| Medical Device Problem Code |
Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor reported that a patient was receiving adjust belt messages.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of electrode belt has been completed.The reported problem (adjust belt messages) has been confirmed.Upon investigation the electrode belt failed an ecg fall-off test.This failure was due to broken wires on the ecg c and d cable.The root cause for the broken wires could not be positively identified.There were no adverse events associated with the damaged belt.
|
| |
|