Model Number FG11815
Medical Device Problem Codes Nonstandard Device (1420); Difficult to Remove (1528); Difficult or Delayed Activation (2577)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/31/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the stent was difficult to deploy.The delivery system also experienced difficulty during removal and exhibited damage.The patient presented as asymptomatic for a right transcarotid artery revascularization (tcar) procedure.An enroute transcarotid stent (tss) system was selected for use.It was reported that the stent was difficult to deploy.Following stent deployment, resistance was encountered during removal of the delivery system.Upon removal, the physician observed that the metal had uncoiled on the delivery system.The procedure was completed, and the patient was reported as stable.No abnormal patient anatomy or lesion was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENROUTE Transcarotid Stent
Common Device Name
Stent, Carotid
Manufacturer (Section D)
Silk Road Medical
1213 innsbruck drive
sunnyvale CA 94089
Manufacturer (Section G)
CORDIS US CORP
14201 nw 60th avenue
miami lakes FL 33014
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
+1(612)875-7234
MDR Report Key50859318
Report Number2124215-2026-50521
Device Sequence Number13138775
Product Code NIM
UDI-Device Identifier00811311021314
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFG11815
Device Catalogue NumberFG11815
Device Lot Number0018544953
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2026
Initial Report FDA Received Date09/22/2026
Date Device Manufactured01/21/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age7 YR
Patient SexMale
-
-