Model Number 12673-03
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/31/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this was an arteriotomy closure of the calcified right common femoral artery using the pre-close technique via an 8f sheath hole prior to an endovascular aneurysm repair (evar) procedure.Reportedly, with 4 proglide devices, the footplates would not close.After removal of the devices, the sheath was upsized to a16f, and the evar procedure was completed.Hemostasis was achieved via cutdown.There were no reported adverse patient effects.No additional information was provided.
 
Additional Manufacturer Narrative
Visual inspection and functional testing were performed on the returned device.The reported difficulty closing the foot was not confirmed during functional testing.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the reported information and the observations from the returned analysis, the investigation determined that the reported issue appears to be related to circumstances of the procedure.In this case, it is likely that challenging anatomical conditions (tissue caught in the foot) led to the reported failure to close the foot.Based on the results of the complaint investigation there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.The additional devices referenced in b5 are filed under separate medwatch report numbers.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Perclose ProGlide¿
Common Device Name
Device, Hemostasis, Vascular
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 -280
Manufacturer (Section G)
ABBOTT VASCULAR, Reg # 3005718570 (P099)
cashel road
clonmel tipperary UNK
EI   UNK
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591--462
+1(951)914-5259
MDR Report Key50859321
Report Number2024168-2026-03657
Device Sequence Number9508758
Product Code MGB
UDI-Device Identifier08717648013089
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P960043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number12673-03
Device Catalogue Number12673-03
Device Lot Number6061041
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/15/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2026
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/10/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexFemale
Patient Weight230 LB
Patient RaceWhite
-
-