The customer reported getting incorrect numbers that are not in range while using the evencare g3 blood glucose test strips and running the evencare g3 glucose control solutions.The glucose monitor was visually inspected before functional testing was conducted, and no signs of physical damage, wear, or defects were observed; functional evaluation was performed using both the customer provided evencare test strips and control solutions, as well as on hand evencare control test strips and control solutions.Based on the sample evaluation, the control solution supplied by the customer was not in range, while the glucose monitor reported as affected functioned as intended as the on hand evencare control solutions were within the specified ranges.It is difficult to determine an exact cause as it is unknown what could have contributed to the issue, and the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|