• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Philips Medical Systems Patient Information Center IX; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Philips Medical Systems Patient Information Center IX; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) Back to Search Results
Model Number Patient Information Center iX
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/11/2026
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Patient Information Center IX
Common Device Name
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Manufacturer (Section D)
Philips Medical Systems
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
Philips Medical Systems
3000 minuteman rd
andover MA 01810
Manufacturer Contact
deborah currlin
3000 minuteman rd
andover, MA 01810
+1(617)245-5900
MDR Report Key50859627
Report Number1218950-2026-100654
Device Sequence Number21320505
Product Code MHX
UDI-Device Identifier00884838093041
UDI-Public(01)00884838093041(10)C.03.10
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K183387
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPatient Information Center iX
Device Catalogue Number866389
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2026
Initial Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/17/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-