| Model Number |
407439 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During the atrial fibrillation procedure, it was noted that the swartz sheath was punctured by the brk needle during transseptal.While initiating transseptal puncture with the brk needle it was noted that there was resistance in the sheath.The entire assembly was withdrawn from the patient, and it was noted that the brk needle had punctured the swartz sheath.The swartz sheath was exchanged, and transseptal puncture was completed successfully.The procedure was completed with no adverse patient health consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The batch record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported puncture and difficult to insert could not be conclusively determined.
|
| |
|