Investigation summary:
the device history record (dhr) for lot 2603509, reference 515629 product set c83-o was reviewed, and no deviations or non-conformances were identified that could be related to the reported event.A review of the applicable risk document indicates that the potential risk of the reported event was assessed appropriately in the risk management documentation.Three retained samples of the c83-o set, lot 2603509, reference 515629, were received for evaluation of the condition reported by the customer.A detailed visual inspection was performed on all three returned samples.No anomalies, damage, or deviations were identified that could be associated with the reported failure mode.As part of the investigation, functional testing was conducted by connecting each of the three retained samples to an n40-o injector.All evaluated units were successfully connected to the luer adapter, achieving proper engagement and locking as intended by the product.No abnormalities or performance issues were observed during testing.
|