| Model Number |
GIF-XP190N |
| Medical Device Problem Codes |
Device Reprocessing Problem (1091); Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the gastrointestinal videoscope for a separate issue.During inspection of the device, it was found that there was no light and there was black foreign material in the air/water nozzle.There was no patient involvement.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
A root cause could not be identified.Based on the results of the investigation, the most likely cause of the no light was an electrical component failure.The most likely cause of the foreign material in the scope was unable to be determined.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the h4 device mfg date.
|
| |
|