| Model Number |
OTV-S200 |
| Medical Device Problem Codes |
Naturally Worn (2988); Data Problem (3196)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the video system center had stopped working in the middle of an unspecified procedure.The procedure was completed using a back-up device.No patient harm was reported.During the initial evaluation of the returned device, the reported event was confirmed and related to the internal buffer being full.Also, a worn-out light guide slider switch was found to have caused an intermittent high intensity mode.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection, and the reported failure was¿confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: a worn-out light guide slider switch that caused an intermittent high intensity mode.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the reported malfunction: a full internal buffer.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|