Catalog Number PN-0006000
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was a low water pressure alarm and burning smell coming from the device during pre-treatment cycle.No patient was involved.
 
Additional Manufacturer Narrative
An outset medical, inc.Field service engineer (fse) evaluated the device on-site, inspected the ultrafilter pump (ufp) and identified damage to ufp during inspection.Therefore, fse replaced the ufp, performed rinse service rinse, and successfully placed console in chemical disinfection.Based on the available evidence, the investigation confirmed thermal damage to the ufp; however, the specific root cause of the failure could not be determined from the available information.A review of production records for this serial number did not note any related manufacturing nonconformances that would contribute to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Tablo Hemodialysis System
Common Device Name
Dialyzer, High Permeability With Or Without Sealed Dialysate System
Manufacturer (Section D)
Outset Medical, Inc.
3052 orchard drive
san jose CA 95134
Manufacturer (Section G)
Outset Medical, Inc.
3052 orchard drive
san jose CA 95134
Manufacturer Contact
dinamarie stefani
3052 orchard drive
san jose, CA 95134
+1(669)231-8200
MDR Report Key50863447
Report Number3010355846-2026-00133
Device Sequence Number9314418
Product Code KDI
UDI-Device Identifier00850001011204
UDI-Public(01)00850001011204(11)230609
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K190793
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPN-0006000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/09/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1