A 72-year-old male patient was undergoing hemodialysis treatment via catheter with treatment initiated at approximately 6:30 am.At approximately 7:44 am, the patient experienced low blood pressure and vomiting, resulting in a code blue being called.Cardiopulmonary resuscitation (cpr) 5-10 compressions was performed as a medical intervention.Following the event, the patient's condition stabilized and the patient was alert.The treating team attributed the event to the patient's preexisting conditions.No alleged tablo malfunction was reported or identified.
|
|
From the information provided, there is no indication that there was any device malfunction, nonconformance, or misuse that contributed to the reported event.Potential adverse events in the instructions for use (ifu) with the tablo system includes, but are not limited to, other, more serious, complications arising from dialysis, such as hemorrhage, air embolism, acidosis, alkalosis or hemolysis, can cause serious patient injury or death.Outset medical, inc.Has reviewed site device logs with a procedure date of (b)(6)2026, no related system alarms were found to have occurred during treatment.Additionally, cross functional review of the treatment logs and event description confirmed normal device operation and did not identify any anomalous system activity that could have contributed to the reported event.Based on the available evidence, no device-related failure was identified.A review of production records for this serial number did not note any related manufacturing nonconformances that would contribute to this event.
|