| Catalog Number |
D134805 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Stroke/CVA (1770)
|
| Date of Event |
07/22/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient underwent an ablation procedure with two (2) thermocool smarttouch sf catheters and the patient experienced a cerebrovascular accident (cva) which required hospitalization and an interventional radiology procedure.It was initially reported there were no errors on the system, and force values were normal.Then the smartablate generator displayed ¿hi¿ force reading when ablation was applied.The catheter was re-zeroed, same problem occurred, replaced the cable but same problem, replaced catheter and the issue resolved issue.The customer's reported force issue is not considered to be mdr reportable since the potential risk that it could cause or contribute to a serious injury or death to the operator or patient is remote.Hours after procedure, the patient became confused and irritable with mental status change.An mri was done which confirmed the cva/stroke.Interventional radiologist did an unknown procedure.Patient required extended hospitalization as a result of the event.Patient's vision has been affected as the patient reported blindness as a result of the cva.The physician¿s opinion on the cause of this adverse event is that it is not related to biosense webster product.However, additional information was not received on what it is related to.On (b)(6) 2026, additional information was received indicating that ablation was performed with the first thermocool smarttouch sf catheter which had the force issue but it was immediately replaced.The physician believes the event was not related to bwi products because the patient may have had calcium in aorta that he believes may have dislodged or ¿showered¿ due to the placement of long sheath in ao that was not bwi product.Based on the additional information that bwi became aware of on 26-aug-2026, the event was reassessed to also be mdr reportable against the first thermocool smarttouch sf.This event had been previously reported under manufacturer report number: 3013300026-2026-02687.
|
| |
|
Additional Manufacturer Narrative
|
|
Device investigation details:
an analysis of the product could not be performed since a physical sample was not received for evaluation.However, if the product is received at a later date, the investigation will be updated as applicable.A manufacturing record evaluation was performed for the finished device number lot 31882491l and no internal actions related to the complaint were found during the review.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref.# (b)(4).
|
| |
|
Search Alerts/Recalls
|
|