| Model Number |
SIGCIRXL |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a robotic-assisted left colectomy, the device was fired and the screen showed the anvil was tilting before removal.After removal, the device was examined and the reload had become detached from the adapter.The reload came off as the integrated trocar extended and prevented the anvil from tilting.After troubleshooting, the reload was able to be returned to the adapter and allowed the anvil to tilt normally.No patient harm or complications.After the case, the adapter and handle were tested with a fresh reload without any issues.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|