A peritoneal dialysis (pd) patient contact reported to fresenius medical care (fme) technical service (ts) on 09/05/2026 that m83 pump a vs.B volume error alarm occurred in fill 1 of 5 and the patient was connected during the incident.The cycler has not been re-setup with new supplies.A fluid leak was discovered during fill 1 of 5.Upon follow-up conducted on 09/10/2026 the patient confirmed the reported event.The patient confirmed they did not experience any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient confirmed treatment was not completed on the day of the reported event.Per patient there were no external leaks from the solution bag or any other connections prior to, during, or after the event.The patient confirmed that there was fluid inside the cycler door.Per the patient they don¿t know what caused the leak since they checked the cassette and the lines and there were no visible damage or holes.The patient stated that they have resumed treatment on the cycler since then with no further issues.The sample is not available to be returned to the manufacturer for physical evaluation.
|