| Model Number |
MR8-AA15 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device was defective.There was no patient impact.Additional information received upon evaluation stating that the bearings were detached.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis: evaluation of the device determined that the malfunction of defective device was confirmed as the bur was not locking into the device.The likely cause of the failure is due to detached bearings.It was also noted that the laser markings were faded.G3: aware date used as date of analysis performed date as the event is reported based on evaluation findings.B3: date of event notification:11.08.2026 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|