It was reported that during a da vinci-assisted surgical procedure, user informed the technical support engineer (tse) that the patient side cart (psc) displayed a message stating that it was running on battery.The tse had the user check the breaker on the back of the psc and the user confirmed it remained in the up position, then had the user verify the system was connected to a known-good outlet that powered other equipment, but the running-on-battery message still displayed.The tse reviewed the system log and noted that the system logs showed the primary power supply operated below load.The user continued with the procedure using backup battery.No known impact or patient consequence was reported.Intuitive surgical, (isi) made multiple follow-up attempts to obtain additional information.However, no further details have been received as of the date of this report.
|
|
An investigation was completed to determine the cause of this reported event.An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.Fse replaced the power supply and the system power manger (spm) to correct the reported event.The system was tested and verified as ready for use.Isi did not receive a da vinci product involved with this complaint to perform failure analysis.The complaint was confirmed based on the field evaluation, which indicates that the device may have contributed to the customer reported issue.
|