| Model Number |
380990-16 |
| Medical Device Problem Code |
Visual Prompts will not Clear (2281)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, an intuitive surgical, inc.(isi) technical support engineer (tse) was contacted after the operating room staff reported that the system displayed error 40006 and that a hard power cycle had not resolved the issue.The site had already engaged field support for troubleshooting prior to contacting the tse, and the tse documented the reported behavior and captured the error and contact identification information for follow-up.The procedure was converted to laparoscopic surgery with no reported injury.Isi followed up with the initial reporter and obtained the following additional information: the procedure was converted to traditional laparoscopic surgery.If converted to traditional laparoscopic, no additional ports were placed.The patient tolerated the change, and the surgery was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.The fse replaced the patient side manipulator (psm).The system was tested and verified as ready for use.Isi did receive a da vinci product involved with this complaint to perform failure analysis.However, failure analysis is not complete.
|
| |
|