| Model Number |
470344-14 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the customer experienced an issue with a curved bipolar dissector instrument.The customer reported event has no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The curved bipolar dissector has been returned and evaluated by the failure analysis (fa) team.The instrument was found to have a dislodged proximal clevis, a broken main tube, and detached fragments of the main tube.The proximal clevis remains attached to the instrument.The main tube was found broken at the distal end, with a fragment detached.The detached fragment was not returned with the instrument.The approximate size of the missing fragment is 1.45 mm x 1.69 mm.
|
| |
|