| Model Number |
480275-10 |
| Medical Device Problem Code |
No Device Output (1435)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, when using the harmonic ace instrument, it was found that it was not recognized and replaced with a new one.The procedure was completed with no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was completed to determine the cause of this reported event.Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The harmonic ace instrument was analyzed and the complaint was not confirmed by failure analysis.The reported event with the instrument could not be replicated nor confirmed.The harmonic ace was tested on an in-house system.The instrument passed the recognition, engagement and initialization tests.The instrument moved intuitively with full range of motion in all directions.The instrument clamp arm opened and closed properly several times.Due to the missing teflon pad the energy delivery test could not be performed.Additional observation not reported by site was that the clamp arm was found to have a missing teflon pad.The complete teflon pad was not returned.Components adjacent to this teflon pad do not show damage.The probable root cause of the teflon pad damage is attributed to use conditions.Damage is caused by heat generation when activating the harmonic ace blade with little to no tissue between the pad and the blade itself.
|
| |
|