| Model Number |
430011-57 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, single port (sp) fenestrated bipolar forceps instrument had misaligned jaws.The procedure was completed with no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The single port (sp) fenestrated bipolar forceps instrument was analyzed and found to have one severely bent grip, causing misalignment of the grip tips.There was a 1.60 mm offset at the tips.The grips were found to have cracking damage.A molded insulator was found to be broken.The instrument was found to have a broken molded insulator on the upper jaw.The broken piece measures approximately 1.37 mm x 1.85 mm, confirming that a fragment detached from the instrument and was not returned with the instrument.The grip base for the grip with the cracked molded insulator does appear to be bent.The complaint regarding misaligned jaws was confirmed by failure analysis.The probable root cause can be attributed to mechanical impact or excessive force applied to the jaws, such as an accidental drop of the instrument.
|
| |
|