| Model Number |
0998-00-0800-65 |
| Medical Device Problem Code |
Failure to Read Input Signal (1581)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the customer that during use the cardiosave intra-aortic balloon pump (iabp) did not display direct ecg input.Patient involved and no harm is reported.Unit was swapped and there was no impact on the patient or the treatment.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that they inspected the device but were unable to reproduce the problem.This time, the ecg cable (green connector) + minisim (simulation) did not receive a signal input.Fse connected the (letter) and performed testing, but they were unable to reproduce the symptoms.Fse also checked the signal input of the external input (moni-x cable), but there were no problems.Pumping operation via direct ecg input is ok.No abnormalities were found in the cardiosave or ecg cables this time.Device is now back in a condition suitable for clinical use.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter: (b)(6).Upon completion of the investigation into this event a follow up report will be submitted.
|
| |
|