| Model Number |
470500-13 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted radical transvesical surgical procedure, when the gas hose was connected, the pressure blew out the inlet pin of the seal; it was then switched to the other seal and it blew out again.The customer used another accessory and proceeded with the case.The procedure was completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical inc.(isi) has not received the da vinci product with an alleged issue to perform failure analysis.A follow-up mdr will be submitted if the product is returned and evaluated and/ or if additional information is received.A review of the provided image was performed.The universal seal pictured showed a broken luer fittings.
|
| |
|
Search Alerts/Recalls
|