C.R. Bard, Inc. (BASD) -3006260740 PowerPort ClearVUE isp Implantable Port, ChronoFlex Single-Lumen, 6F; Port & Catheter, Implanted, Subcutaneous, Intravascular Back to Search Results
Catalog Number 1606062
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6)2026, a patient underwent a port placement procedure using the powerport clearvue isp.Post procedure, it was reported that an x ray was performed to assess for complications and confirmed that the ct marker was inverted; however, the cv port itself was not inverted.A review of 50 cases implanted during august identified one additional case with a similarly inverted marker.It was further noted that the issue was initially identified in a batch manufactured on august 27, and a retrospective review revealed the same issue in a batch manufactured on august 14.There was no reported patient injury.
 
Additional Manufacturer Narrative
H11: as the lot number for the device was provided, a review of the device history records will be performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.Section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PowerPort ClearVUE isp Implantable Port, ChronoFlex Single-Lumen, 6F
Common Device Name
Port & Catheter, Implanted, Subcutaneous, Intravascular
Manufacturer (Section D)
C.R. Bard, Inc. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
Bard Reynosa S.A. de C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
michael gaffney
800 w. rio salado pkwy
tempe, AZ 85281
+1(480)303-2689
MDR Report Key50865152
Report Number3006260740-2026-06914
Device Sequence Number9425321
Product Code LJT
UDI-Device Identifier00801741026416
UDI-Public(01)00801741026416(17)271130(10)RELN0704
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K063377
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1606062
Device Lot NumberRELN0704
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-